Version:
Portable gas exchange fluorescence system
ca. 40 ml, 3 cm x 5 cm x 2.7 cm (L x W x H)
Adapter IMAG‐MAX/GWK1 for IMAGING-PAM (MAXI-Version) on Gas Exchange Chamber (3010-GWK1)
Adapter Plate with legs and eye protection for positioning IMAG-MAXI Head on 3010-GWK1
2 Aluminum boxes with individual foam lining for GFS-3000 and accessories
60 cm x 40 cm x 35 cm (L x W x H)
5 kg (each)
2 Aluminum boxes with individual foam lining for GFS-3000 and accessories
60 cm x 40 cm x 35 cm (L x W x H)
5 kg (each)
LED-Panel fitting to the gas exchange chamber 3010-GWK1: Illuminated area 14 cm x 12 cm.
LED Colors: red, green, blue and white; maximum output (all colors together): 2000 μmol m-2 s-1 or better.
Air cooled; with separate power supply. Colors can be mixed with red, green, blue, and white in steps of intensity 0 to 100 or more; total intensity can be chosen in steps of 0.1%; operation with GFS-Win software via RS485 connector.
Blue: 455 ±10 nm, HBW (half bandwidth) 440 - 460 nm
Green: 525 ±10 nm, HBW 500 - 545 nm
Red: 625 ±5 nm, HBW 620 - 640 nm
White: 450 ±10 nm, HBW 435 - 460 nm; second peak at 590 ±25 nm, HBW 510 - 640 nm with tail up to 800 nm
± 10% within the 14x12 cm area or ± 3 μmol m-2 s-1 (whatever is bigger).
PAR sensors with multiplier between -50 and -800 μmol m-2 s-1 per μA result in a range of 0 to 6400 μmol m-2 s-1, resolution: 1 μmol m-2 s-1
10 A slow-blow fuse, 5x20 mm
16 V, 8 A
AC Power Supply 3020-N for laboratory operation
± 10% within the 14x12 cm area or ± 3 μmol m-2 s-1 (whatever is bigger).
27 cm x 19 cm x 13 cm L x W x H
Adapter IMAG‐MIN/GFS for IMAGING-PAM (MINI-Version) on Gas Exchange Standard Measuring Head (3010-S)